Metrohm tiamo 1.0 Manual User Manual
Page 461

Commands
437
Command name
25 characters, [ MEAS T/Flow # ]
Name of the command.
The general parameters for the control device are defined on this tab.
Device
Device name
Device name, [ not defined ]
Selection of a 774 Oven Sample Processor from the device table. Addition-
ally,
not defined
can be selected. In this case, the device name must be as-
signed by the user at the start of the method.
Device type
774 Oven Sample Processor
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.
If the option
not defined
is selected as
Device name
, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Measuring parameters
Command name
25 characters, [ MEAS T/Flow # ]
Name of the command.
The measuring parameters are defined on this tab.
Measuring time
0 ... [ 600 ] ... 999'999 s
Maximum period of time for measurement.
Measuring interval
1 ... [ 10 ] ... 999'999.0 s
Time interval for entering a measuring point in the list of measuring points.
Stop measured value Temp.
-20.0 ... 150.0 °C, [ off ]
Stop when the preset measured value has been reached since the start of
the measurement.
off
means no stop.