Determination parameters – Metrohm tiamo 1.0 Manual User Manual
Page 143

Subwindow Run
119
Determination parameters
User
50 characters
If the option
Enforce login with user name
is switched on in the Security set-
tings then the short name of the user currently logged in will be shown in
this field. The field cannot then be edited. If work is carried out without log
in then a user name can be entered in this field (only in the status
READY
).
Sample number
[ 0 ] ... 99999
At each program start the sample number will be set to
0
for all workplaces.
Each time that a determination is started it will be increased by
+1
. It can
only be edited by the user in the status
READY
.
Remark
100 characters
Freely definable remark about the determination that is saved together with
the determination. Instead of a manual entry it is also possible to select a
prepared Text template.
Note
The field
Remark
can also be edited live while determinations are being carried
out. With the context-sensitive menu item
Modify remark
the
Live modifications
window opens.
Autostart
1 ... 999, [ Sample table ]
Number of automatic, internal starts of determinations within a sample se-
ries.
The first field after
Autostart
shows the number of determinations that have
already been carried out (actual counter). This field cannot be edited, but in
the status
READY
or
PAUSE
the context-sensitive menu item
Reset autostart
counter
can be used to reset it to
0
.
The second field after
Autostart
shows the total number of determinations to
be carried out per series (set counter). It contains the entry
Sample table
as
standard which indicates the number of lines of the working sample table.
This field can only be edited in the status
READY
.
Note
The second field after
Autostart
can also be altered live while determinations are
being carried out. With the context-sensitive menu item
Modify autostart counter
the window
Live modifications
opens.
Statistics
[ on ], off
This field is only shown when a method has been loaded in whose
START
command the option
Statistics
is switched on. The statistics defined in the
method can be switched on and off manually here (only in the status
READY
).
The first field after
Statistics
shows the number of determinations that have
already been carried out for the statistics (actual counter). This field cannot
be edited, but in the status
READY
the context-sensitive menu item
Delete
statistics
can be used to reset it to
0
, which at the same time deletes the
statistics data that has already been calculated.
The second field after
Statistics
shows the number of determinations to be
carried out for the statistics (set counter). It contains the value for the
Num-
ber of single determinations
defined in the
START
command as standard.
This field can only be edited in the status
READY
.