Metrohm tiamo 1.0 Manual User Manual
Page 611

Subwindow Devices
587
New version
Shows the new device program version to be loaded.
Loads new program version.
Note
Make sure that the device is not manipulated or switched off during the loading
process and follow the instructions shown.
MSB #
Request for dosing device preparation
Selection when the request for carrying out the command
PREP
(prepare) is to be
shown for the dosing device connected to the MSB.
At program start
[ on ], off
If this option is switched on then at each program start the request to pre-
pare the dosing device will appear.
On attaching an exchange/dosing unit
[ on ], off
If this option is switched on then each time that an exchange unit is at-
tached to the Dosimat or each time that a dosing unit is attached to the
Dosino the request to prepare the dosing device will appear.
Time interval
on, [ off ]
If this option is switched on then the request to prepare the dosing device
will appear after the time interval defined here.
all
0.1 ... [ 12.0 ] ... 999.9 h
Time interval after which the request to prepare the dosing device will ap-
pear.
Connected devices
Information about the following peripheral devices connected to the MSB connec-
tion appears here:
•
Dosing device
The
Dosing device type
is shown here for the connected dosing de-