Commands

397

If the EP has been reached, after the last dosing the delay time here de-
fined is awaited and then the titration is stopped. This parameter is only

shown for

Stop criterion

=

time

.

inf.

stands for infinite.

Stop time

1 ... 999999 s

(only Titrino 720, 736, 751, 758, 784, 795, 799)

The titration is stopped if the defined stop time has elapsed. This parameter
is only shown for

Stop criterion

=

time

and

Delay time

=

inf.

.

Titration parameters

Command name

25 characters, [ KFT Ipol # ]

Name of the command.

Parameters defining the run of the titration.

Titration direction

Selection of the titration direction. If two endpoints are set the direction is al-
ready given and the following selection is ignored.

+

positive measured value alteration

[ - ]

negative measured value alteration

auto

The titration direction is determined automatically from the start measured
value and the set endpoint.

Extraction time

[ 0 ] ... 999999 s

The titration is not stopped until the extraction time has elapsed (even when
the EP has already been reached). For example, the entry of an extraction
time is a good idea for the titration of sparingly soluble samples.

Temperature

-20.0 ... [ 25.0 ] ... 150.0 °C

(Titrando)

-170.0 ... [ 25.0 ] ... 500.0 °C

(Titrino)

Manually entered titration temperature. If a temperature sensor is con-
nected and the

Temperature measurement

in the tab

General/Hardware

under

Sensor

is set to

automatic

or

continuous

then the temperature will be meas-

ured continuously.

Time interval measuring point

0.1 ... [ 2.0 ] ... 999999 s

(Titrando)

0.08 ... [ 2.00 ] ... 16200 s

(Titrino)

Time interval for entering a measuring point in the list of measuring points.

Stop conditions

Screenshot and parameters: see SET pH - Stop conditions