Metrohm tiamo 1.0 Manual User Manual
Page 421

Commands
397
If the EP has been reached, after the last dosing the delay time here de-
fined is awaited and then the titration is stopped. This parameter is only
shown for
Stop criterion
=
time
.
inf.
stands for infinite.
Stop time
1 ... 999999 s
(only Titrino 720, 736, 751, 758, 784, 795, 799)
The titration is stopped if the defined stop time has elapsed. This parameter
is only shown for
Stop criterion
=
time
and
Delay time
=
inf.
.
Titration parameters
Command name
25 characters, [ KFT Ipol # ]
Name of the command.
Parameters defining the run of the titration.
Titration direction
Selection of the titration direction. If two endpoints are set the direction is al-
ready given and the following selection is ignored.
+
positive measured value alteration
[ - ]
negative measured value alteration
auto
The titration direction is determined automatically from the start measured
value and the set endpoint.
Extraction time
[ 0 ] ... 999999 s
The titration is not stopped until the extraction time has elapsed (even when
the EP has already been reached). For example, the entry of an extraction
time is a good idea for the titration of sparingly soluble samples.
Temperature
-20.0 ... [ 25.0 ] ... 150.0 °C
(Titrando)
-170.0 ... [ 25.0 ] ... 500.0 °C
(Titrino)
Manually entered titration temperature. If a temperature sensor is con-
nected and the
Temperature measurement
in the tab
General/Hardware
under
Sensor
is set to
automatic
or
continuous
then the temperature will be meas-
ured continuously.
Time interval measuring point
0.1 ... [ 2.0 ] ... 999999 s
(Titrando)
0.08 ... [ 2.00 ] ... 16200 s
(Titrino)
Time interval for entering a measuring point in the list of measuring points.
Stop conditions
Screenshot and parameters: see SET pH - Stop conditions