Metrohm tiamo 1.0 Manual User Manual
Page 331

Commands
307
Number of single determinations
1 ... 99, [ not defined ]
The number of single determinations defined here is valid for all
CALC
commands in the method. When
not defined
is selected, the number of sin-
gle determinations must be defined in the Workplace.
Conditioning
Automatic conditioning
[ on ], off
If this parameter is enabled, the conditioning will be started automatically
at the start of a determination for the all commands with activated condi-
tioning (
SET
,
KFT
,
KFC
). After the track including the command with condi-
tioning has been finished, the conditioning is started again and will be con-
tinued even at the end of the determination.
If this parameter is disabled, the conditioning won't be started until the start
of the command with the conditioning. The conditioning isn't started again
after the track including the command with conditioning has been finished.
This parameter must be switched off for automation applications.
Application note
Command name
25 characters, [ Main track ]
Name of the command.
The application note defined here is automatically displayed in the live window of
the workplace when loading this method.
The Text editor to enter or change the application note is opened with
or by
double-clicking into the text field.