Commands

307

Number of single determinations

1 ... 99, [ not defined ]

The number of single determinations defined here is valid for all

CALC

commands in the method. When

not defined

is selected, the number of sin-

gle determinations must be defined in the Workplace.

Conditioning

Automatic conditioning

[ on ], off

If this parameter is enabled, the conditioning will be started automatically
at the start of a determination for the all commands with activated condi-

tioning (

SET

,

KFT

,

KFC

). After the track including the command with condi-

tioning has been finished, the conditioning is started again and will be con-
tinued even at the end of the determination.
If this parameter is disabled, the conditioning won't be started until the start
of the command with the conditioning. The conditioning isn't started again

after the track including the command with conditioning has been finished.
This parameter must be switched off for automation applications.

Application note

Command name

25 characters, [ Main track ]

Name of the command.

The application note defined here is automatically displayed in the live window of
the workplace when loading this method.

The Text editor to enter or change the application note is opened with

or by