Page 612

background image

588

Configuration

vices; for type 8xx dosing devices the

Dosing device serial number

is

also shown.

Stirrer

The

Stirrer type

and

Stirrer serial number

are shown here for the con-

nected stirrers.

Remote box

No properties can be shown for a connected remote box.

GLP

GLP test date

Date selection

Date of the last GLP test. This date can be selected after pressing

in the

dialog window

Select date

(see General program functions - Edit - Select

date).

Comment on GLP test

1000 characters, [ empty ]

Comments about GLP test.

Monitoring of GLP validity

on, [ off ]

If this option is switched on then the time interval for the GLP test will be
monitored.

Note
This option can only be switched on when a date has been entered under

GLP

test date

.

GLP test interval

0 ... [ 999 ] days

Time interval to next GLP test. If an interval is entered here then the date for
the

Next GLP test

will be adapted automatically.

Next GLP test

Date selection, [GLP test date + 999 days ]

Date on which the next GLP test is to be carried out. This date can be se-
lected after pressing

in the dialog window

Select date

(see General

program functions - Edit - Select date). After the date has been entered the
value for the

GLP test interval

will be adapted automatically.