362

Method

Min. waiting time

[ 0 (slow, optimal, fast) ] ... 999999 s

(only Titrando)

The minimum waiting period is only important for drift-controlled measure-
ments. Measured value acceptance only takes place when the minimum
waiting time has elapsed, even when the measured value drift has already
been achieved. The drift continues to be checked while the waiting time is
elapsing.

Max. waiting time

0 ... [ 38 (slow) ] ... [ 26 (optimal) ] ... [ 21 (fast) ] ... 999999 s

(Titrando)

0 ... [ 38 (slow) ] ... [ 26 (optimal) ] ... [ 21 (fast) ] ... 9999 s

(Titrino)

If measured value drift is switched off or has not yet been achieved then the
measured value will be accepted after the maximum waiting time has
elapsed. If the waiting time has not been newly entered then a waiting time

that is suitable for the drift will be calculated automatically according to the
following equation:

Dosing of increments

(only visible for

Titration rate

=

user

)

Volume increment

0.0001 ... [ 0.05 (slow) ] ... [ 0.10 (optimal) ] ... [ 0.20 (fast) ] ... 9.9999 mL

(Ti-

trando),

0.00 ... [ 0.05 (slow) ] ... [ 0.10 (optimal) ] ... [ 0.20 (fast) ] ... 9.999 mL

(Titrino)

Small volume increments are used for the determination of blank values or
for very asymmetrical curves. A good guideline is 1/20 of the expected EP
volume. For steep jumps the volume increment should be nearer 1/100 of
the expected EP volume and for flat jumps nearer 1/10.

Dosing rate

0.01 ... 166.00 mL/min, [ maximum ]

(Titrando)

0.01 ... 150.00 mL/min, [ maximum ]

(Titrino)

Speed at which the volume increments are added. The maximum dosing
rate depends on the cylinder volume of the used exchange unit or dosing
unit.

Temperature

Temperature

-20.0 ... [ 25.0 ] ... 150.0 °C

(Titrando)

-170.0 ... [ 25.0 ] ... 500.0 °C

(Titrino)

Manually entered titration temperature. If a temperature sensor is con-
nected and the

Temperature measurement

in the tab

General / Hardware

un-

der

Sensor

is set to

automatic

or

continuous

then the temperature will be

measured continuously. This value is used for temperature correction in pH
measurements.

Stop conditions

Screenshot and parameters: see DET pH - Stop conditions

Potentiometric evaluation

Screenshot and parameters: see DET pH - Potentiometric evaluation