7 remote functions – Metrohm tiamo 1.0 Manual User Manual
Page 93

Manual control
69
2.6.7 Remote
functions
If in the subwindow for Select device a remote box or remote interface is selected
then in the subwindow for Functions/Parameters the functions and parameters
that are possible for the device will appear.
Devices
The remote functions can be carried out with the following instruments:
Titrino: 751, 758, 784, 785, 795, 798, 799
Titrando: 808, 809, 835, 836, 841, 846
Coulometer: 756, 831
Sample Changer: 730, 774, 778, 789, 814, 815
Inputs
Current status
Shows the current status of the 8 input lines.
Templates/Entry
If the current status corresponds to one of the defined Templates for the in-
puts then the corresponding name will be shown here.
Outputs
Current status
Shows the current status of the 14 output lines.
Templates/Entry
Binary pattern consisting of exactly 14 characters (0, 1, *, p), [
************** ], signal template
Input of the binary pattern for the output signal of exactly 14 characters or
selects a predefined Signal template.
The following characters can be entered:
0
= line inactive,
1
= line active,
*
= any line status
p
= set impulse (impulse length = 200 ms. If the output is to be an impulse
with a different length then an appropriate template must be defined for it.)
The output lines and bits are numbered from right to left:
Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0
Bit
13 12 11 10 9 8 7 6 5 4 3 2 1 0