Metrohm tiamo 1.0 Manual User Manual

Page 394

background image

370

Method

Additional measured values

Definition of additional measured values of other measuring com-
mands which are saved as additional column in the list of measuring

points.

General/Hardware

Screenshot and parameters: see DET pH - General/Hardware

Start conditions

Command name

25 characters, [ SET pH # ]

Name of the command.

The start conditions are processed in the listed sequence before the titration is
started.

Initial measured value

The initial measured value is determined before the other start conditions are
processed. The measured value acceptance depends on the three following pa-
rameters:

Signal drift

0.1 ... 999.0 mV/min, [ off ]

(only Titrando)

The measured value is only accepted if the drift undercuts the entered
value.

off

means that the measured value is accepted after the maximum

waiting time.

Min. waiting time

[ 0 ] ... 999999 s

(only Titrando)

The minimum waiting time is only important for drift-controlled measure-
ments. The measured value is accepted only after the minimum waiting
time has elapsed, even if the signal drift has already been achieved. The
drift continues to be checked as the minimum waiting time is passing.