Metrohm tiamo 1.0 Manual User Manual
Page 394

370
Method
•
Additional measured values
Definition of additional measured values of other measuring com-
mands which are saved as additional column in the list of measuring
points.
General/Hardware
Screenshot and parameters: see DET pH - General/Hardware
Start conditions
Command name
25 characters, [ SET pH # ]
Name of the command.
The start conditions are processed in the listed sequence before the titration is
started.
Initial measured value
The initial measured value is determined before the other start conditions are
processed. The measured value acceptance depends on the three following pa-
rameters:
Signal drift
0.1 ... 999.0 mV/min, [ off ]
(only Titrando)
The measured value is only accepted if the drift undercuts the entered
value.
off
means that the measured value is accepted after the maximum
waiting time.
Min. waiting time
[ 0 ] ... 999999 s
(only Titrando)
The minimum waiting time is only important for drift-controlled measure-
ments. The measured value is accepted only after the minimum waiting
time has elapsed, even if the signal drift has already been achieved. The
drift continues to be checked as the minimum waiting time is passing.