Commands

405

Extraction time

[ 0 ] ... 999999 s

The titration is not stopped until the extraction time has elapsed (even when

the EP has already been reached). For example, the entry of an extraction
time is a good idea for the titration of sparingly soluble samples.

Temperature

-170.0 ... [ 25.0 ] ... 500.0 °C

Manually entered titration temperature.

Time interval measuring point

0.4 ... [ 4.0 ] ... 16200 s

Time interval for entering a measuring point in the list of measuring points.

Stop conditions

Command name

25 characters, [ KFC # ]

Name of the command.

Condition for stopping the titration.

Stop time

0 ... 999999 s, [ off ]

Stops when the preset time has elapsed since the start of the titration.

off

means no stop.

Conditioning

Command name

25 characters, [ KFC # ]

Name of the command.

Conditioning

[ on ], off

Switching conditioning on/off. If conditioning is deactivated all following pa-
rameters are not visible.