122

Workplace

Check solutions

Checks whether the solutions demanded in the device-specific com-

mands are present and available.

Check sensors

Checks whether the sensors demanded in the device-specific com-
mands are present and available.

Check sample data

Checks whether the entered sample data corresponds to that defined

in method variables and is valid.

Note
The Start test for the number of lines defined under

Autostart

can also be trigge-

red manually in the

READY

status with the menu items

Tools, Run test

or

Sample

table, Run test

.

4 - Preconditioning (only for methods with conditioning)

In methods that contain commands with switched-on conditioning and in which

the option

Automatic conditioning

is switched on in the

START

command then pre-

conditioning will start for all these commands. Preconditioning does not belong to
the determination itself and does not increase any counter (

Sample number

or

Sta-

tistics

).

Conditioning requirement not fulfilled

After the start of preconditioning the status changes to

COND BUSY

. In

this status conditioning can be interrupted with

[Hold]

and continued

with

[Cont]

. If an error occurs during conditioning then neither the Error

track nor the Exit track will be carried out.

Conditioning requirement fulfilled

If the conditioning requirement has been fulfilled the status changes to

COND READY

and the method then starts automatically.

5 - Start method

At the start of the method the following checks and actions are triggered:

Assign determination ID

An unambiguous determination ID is assigned for the determination.

Reserve device (module)

The devices (or their functional units) used by the device-dependent
commands contained in the method are reserved for the duration of
the determination, i.e. they cannot be used in other workplaces nor
can they be operated manually.

Increment sample number

The

Sample number

is increased by

+1

.

Increment statistics counter

If statistics has been switched on in both the

START

command and on

the tab

Determination series

then the statistics counter will be increased

by

+1

. If a new method is loaded or if the statistics counter agrees with

the set counter then the current statistical data will first be deleted and
the statistics counter reset to

0

.