420

Method

Stop measured value

-2'000.0 ... 2'000.0 mV, [ off ]

(Titrando only)

Stop when the preset measured value has been reached since the start of
the measurement.

off

means no stop.

Temperature

Temperature

-20.0 ... [ 25.0 ] ... 150 °C

(Titrando)

-170.0 ... [ 25.0 ] ... 500 °C

(Titrino)

Manually entered measuring temperature. If a temperature sensor is con-

nected and

temperature measurement

on the tab General/Hardware is set to

automatic

or

continuous

, the temperature is measured continuously. The

value is used for the temperature adjustment during the measurement.

Evaluations

Command name

25 characters, [ MEAS U # ]

Name of the command.

The following methods for evaluating measuring curves are defined on this tab.

Fix end point evaluation

on, [ off ]

To a fixed quantity (measured value or time) the associated values for the

other quantities are interpolated from the list of measuring points for the
fixed end point.

Fixed quantity

[ Measured value ], Time

Selection of the fixed quantity to which the associated values for the other
quantities are to be interpolated from the list of measuring points.

Fix EP# at

-2'000.0 ... 2'000.0 mV, [ off ]

(Titrando)

-2'000 ... 2'000 mV, [ off ]

(Titrino)

Fixed end point 1 ... 9 for

Fixed quantity

=

Measured value

.