Metrohm tiamo 1.0 Manual User Manual
Page 444

420
Method
Stop measured value
-2'000.0 ... 2'000.0 mV, [ off ]
(Titrando only)
Stop when the preset measured value has been reached since the start of
the measurement.
off
means no stop.
Temperature
Temperature
-20.0 ... [ 25.0 ] ... 150 °C
(Titrando)
-170.0 ... [ 25.0 ] ... 500 °C
(Titrino)
Manually entered measuring temperature. If a temperature sensor is con-
nected and
temperature measurement
on the tab General/Hardware is set to
automatic
or
continuous
, the temperature is measured continuously. The
value is used for the temperature adjustment during the measurement.
Evaluations
Command name
25 characters, [ MEAS U # ]
Name of the command.
The following methods for evaluating measuring curves are defined on this tab.
Fix end point evaluation
on, [ off ]
To a fixed quantity (measured value or time) the associated values for the
other quantities are interpolated from the list of measuring points for the
fixed end point.
Fixed quantity
[ Measured value ], Time
Selection of the fixed quantity to which the associated values for the other
quantities are to be interpolated from the list of measuring points.
Fix EP# at
-2'000.0 ... 2'000.0 mV, [ off ]
(Titrando)
-2'000 ... 2'000 mV, [ off ]
(Titrino)
Fixed end point 1 ... 9 for
Fixed quantity
=
Measured value
.