Metrohm tiamo 1.0 Manual User Manual
Page 366

342
Method
General/Hardware
Command name
25 characters, [ DET Ipol # ]
Name of the command.
The general parameters for the control device, the dosing device, sensor and the
stirrer are defined on this tab.
Device
Device name
Device name, [ not defined ]
Selection of the device from device table. Only those devices being able to
execute the command are selectable. Additionally,
not defined
can be se-
lected. In this case, the device name must be assigned by the user at the
start of the method.
Device type
Device types, [ Titrando ]
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.
If the option
not defined
is selected as
Device name
, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Dosing device
Dosing device
[ 1 ] ... 4
(Titrando)
[ internal D0 ]
(Titrino)
external D1, external D2
(only Titrino 736, 751, 758, 799)
Selection of the number of the dosing device (dosing or exchange unit) to
be used for the dosing.