Page 404

background image

380

Method

Titration rate

Titration rate

Three predefined sets of parameters can be selected for the titration rate:

slow

,

optimal

and

fast

. For these sets the parameters for the

Control

are not

displayed. In order to edit these parameters the value

user

must be se-

lected.

slow

For titrations in which the finest details are also to be visible. However, this
can also increase noise and lead to the production of unwanted EPs.

[ optimal ]

Parameter set for all standard titrations; optimized for the most frequent
applications.

fast

For less critical rapid titrations.

user

Editing the individual titration parameters which affect the titration rate.

Control

(only visible for

Titration rate

=

user

)

Dynamics

0.1 ... [ 300.0 (slow) ] ... [ 100.0 (optimal) ] ... [ 30.0 (fast) ] ... 2000.0 mV, off

(Titrando),

1 ... [ 300 (slow) ] ... [ 100 (optimal) ] ... [ 30 (fast) ] ... 2000 mV, off

(Titrino)

Dynamics defines the measured value range before the given endpoint.
This parameter has an important influence on the titration rate and with it on
the accuracy. Within the control range the steps which are added are con-
trolled by the

Min. rate

. The nearer the endpoint the slower the dosing takes

place until the

Min. rate

has been reached. Outside the range the dosing

takes place with

Max. rate

(see Control range).

Max. rate

0.01 ... [ 1.00 (slow) ] ... [ 10.00 (optimal) ] ... 166.00 mL/min, [ maximal (fast) ]

(Titrando)

0.01 ... [ 1.00 (slow) ] ... [ 10.00 (optimal) ] ... 150.00 mL/min, [ maximal (fast) ]

(Titrino)
Speed of addition outside the control range. The maximum dosing speed

depends on the cylinder volume of the exchange unit or dosing unit used.

Min. rate

0.01 ... [ 5.00 (slow)] ... [ 25.00 (optimal) ] ... [ 50.00 (fast) ] ... 9999.00 µL/min

Speed of addition at the start and finish of the titration. The slower the
minimum rate that is selected, the slower the titration and the higher the ac-
curacy.

Stop criterion

Stop criterion

[ drift ], time, off

The titration stops when the endpoint has been reached and the stop crite-
rion has been fulfilled. The titration can be stopped when a particular

drift

has been achieved or after a preset

time

. If no stop criterion has been se-

lected the titration will not be stopped.

Note
The Stop conditions will always stop the titration even if the stop criterion has not
been fulfilled or has been turned off.