Method

Commands

Depending on the measured value one of the following two KFT commands can

be chosen:

KFT Ipol

Voltametric measurement with selectable polarization current (meas-
ured value voltage U).

KFT Upol

Amperometric measurement with selectable polarization voltage
(measured value current I).

KFT Ipol

Command for coulometric Karl Fischer Titrations with voltametric measurement
(selectable polarization current).

Devices

This command can be executed with the following devices which have the KFT

mode:

Titrino: 720, 736, 751, 758, 784, 795, 799

Titrando: 835, 836, 841

Appearance

The command has the following appearance:
without conditioning with conditioning

Parameters

The parameters for the command

KFT Ipol

are configured in the following 8 tabs:

General/Hardware

Parameters for devices, dosing devices, sensors and stirrers.

Start conditions

Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-

tion.

Control parameters

Control parameter for the endpoint (measured value for the endpoint,
titration rate, user defined parameters, stop criterion).

Titration parameters

Parameters for the run of the titration.

Stop conditions

Definitions of conditions which cause the stop of the titration.

Conditioning

Switching conditioning on/off. Definition of start drift, drift correction

and stop conditions.

Additional evaluations

Definition of additional methods for evaluation of the titrations curves.