Metrohm tiamo 1.0 Manual User Manual
Page 680

656
Configuration
Warning limits for slope
on, [ off ]
If this option is switched on then the values for the slope will be shown in
orange if the limits are infringed.
Lower limits
-2.147E+9 ... 2.147E+9
pH electrode:
97.5 %
ISE (+1):
57.2 mV
ISE (-1):
-61.2 mV
ISE (+2):
27.7 mV
ISE (-2):
-31.7 mV
ISE (+3):
17.7 mV
ISE (-3):
-21.7 mV
ISE (+4):
12.8 mV
ISE (-4):
-16.8 mV
Lower warning limits for the slope.
Upper limits
-2.147E+9 ... 2.147E+9
pH electrode:
102.5 %
ISE (+1):
61.2 mV
ISE (-1):
-57.2 mV
ISE (+2):
31.7 mV
ISE (-2):
-27.7 mV
ISE (+3):
21.7 mV
ISE (-3):
-17.7 mV
ISE (+4):
16.8 mV
ISE (-4):
-12.8 mV
Upper warning limits for the slope.
Intervention limits for the slope
on, [ off ]
If this option is switched on then the values for the slope will be shown in
red if the limits are infringed.
Lower limits
-2.147E+9 ... 2.147E+9
pH:
95.0 %
ISE (+1):
55.0 mV
ISE (-1):
-62.4 mV
ISE (+2):
26.7 mV
ISE (-2):
-32.7 mV
ISE (+3):
16.7 mV
ISE (-3):
-22.7 mV
ISE (+4):
11.8 mV
ISE (-4):
-17.8 mV
Lower intervention limits for the slope.
Upper limits
-2.147E+9 ... 2.147E+9
pH:
105.0 %
ISE (+1):
62.4 mV
ISE (-1):
-55.0 mV
ISE (+2):
32.7 mV
ISE (-2):
-26.7 mV
ISE (+3):
22.7 mV
ISE (-3):
-16.7 mV
ISE (+4):
17.8 mV
ISE (-4):
-11.8 mV
Upper intervention limits for the slope.
Warning limits for pH(0) and E(0)
on, [ off ]
If this option is switched on then the values for pH(0) or E(0) will be shown
in orange if the limits are infringed.