Metrohm tiamo 1.0 Manual User Manual

Page 662

background image

638

Configuration

Cylinder volume

1, 5, 10, [ 20 ], 50 mL

Cylinder volume of exchange unit. With intelligent exchange units the cylin-

der volume is read off automatically and cannot be edited. If in the method
you have selected a solution in a titration or dosing command then the cyl-
inder volume will be checked in the sequence.

Cylinder serial number

24 characters, [ 'empty' ]

Serial number of cylinder. It is printed on new cylinders. With intelligent ex-

change units it is read off automatically. It can be altered at any time, e.g. if
the cylinder has to be replaced.

Parameters for preparation

Configures the parameters to be used in the command

PREP

.

Volume

0.00000 ... 99999.99999 mL, [ cylinder volume ]

Volume to be dosed in during preparation. With

Cylinder volume

the whole

contents of the cylinder will be dosed in.

Cycles

1 ... [ 2 ] ... 9

Number of rinsing cycles during preparation.

Dosing rate

0.01 ... 150.00 mL/min, [ maximum ]

Speed at which the solution is to be dosed. The maximum dosing rate de-
pends on the cylinder volume of the exchange unit used (see below). When

the function is carried out the speed will automatically be reduced to the
largest possible value.

Filling rate

0.01 ... 150.00 mL/min, [ maximum ]

Speed at which the cylinder is to be filled. The maximum filling rate de-
pends on the cylinder volume of the exchange unit used (see below). When

the function is carried out the speed will automatically be reduced to the
largest possible value.

Maximum Dosing/Filling rate for exchange unit as a function of the cylinder volume:

Cylinder volume Maximum rate

1 mL

3.0 mL/min

5 mL

15.0 mL/min

10 mL

30.0 mL/min

20 mL

60.0 mL/min

50 mL

150.0 mL/min

Note
Enter lower rates for high-viscosity liquids.

Tubing parameters

Defines the length and diameter of the tubing connected to the exchange unit.
These parameters are important for carrying out the exchange unit command

PREP

correctly, as the volumes of the tubing connections have to be taken into

account.