Cal loop conc – Metrohm tiamo 1.0 Manual User Manual
Page 477

Commands
453
Temperature
Temperature
-20.0 ... [ 25.0 ] ... 150 °C
Manually entered measuring temperature. If a temperature sensor is con-
nected and
temperature measurement
on the tab General/Hardware is set to
automatic
or
continuous
, the temperature is measured continuously. The
value is used for the temperature adjustment during the measurement.
CAL LOOP Conc
Loop command for calibrating ion-selective electrodes. The command is used
together with the command
CAL MEAS Conc
.
Appearance
The command has the following appearance:
Parameters
The parameters for the command
CAL LOOP Conc
are configured in the following
dialog window:
•
CAL LOOP Conc
CAL LOOP Conc - properties
The number of standards and their concentration are defined in the dialog window
CAL LOOP Conc
.
Command name
25 characters, [ Calibration loop Conc # ]
Name of the command.
Standards
Number of standards
1 ... [ 2 ] ... 5
Selection of the number of standards for the concentration calibration.
Unit conc.
mol/L, mmol/L, [ ppm ], %, g/L, mg/L, mg/mL, µg/L
Selection of the concentration unit of the standards.
Conc. standard #
0.001 ... 999'999.999
Entry of the corresponding concentration of each standard.