Metrohm tiamo 1.0 Manual User Manual
Page 451

Commands
427
Stirrer
Stirrer
[ 1 ] ... 4, off
(Titrando only)
Selection of the stirrer.
off
means that no stirrer will be used.
Stirring rate
-15 ... [ 8 ] ... 15
(Titrando only)
Setting the stirrer speed. The direction in which stirring takes place changes
as the sign in front of the stirrer speed changes.
Switch off automatically
[ on ], off
(Titrando only)
The stirrer is automatically switched off at the end of the command if this
option is activated.
Measuring parameters
Command name
25 characters, [ MEAS Upol # ]
Name of the command.
The measuring parameters are defined on this tab.
Measurement
The two measuring modes Measurement with drift control or Meas-
urement without drift control are selectable.
Note
A constant measured value is often only achieved after a certain time, as mixing
and possibly the reaction itself require a certain time. The response time of an
electrode can also increase with time, i.e. reaching a constant measured value
takes longer and longer. In such a case
drift-controlled measurement is par-
ticularly advisable, as the measured values are only accepted when equilibrium
has almost been achieved.
Measurement with drift control
[ on ], off
The measurement is carried out drift-controlled if this option is activated.
The measurement is stopped as soon as the
Signal drift
or the
Stop meas-
ured value
is reached or a
waiting time
has elapsed. The corresponding pa-
rameters are not visible if this option is deactivated.