Commands

403

Control parameters

Command name

25 characters, [ KFC # ]

Name of the command.

Endpoint

EP at

-2000.0 ... [ 50.0 ] ... 2000.0 mV

Measured value for the end point.

Control

Dynamics

0 ... [ 70 ] ... 2000 mV, off

The dynamics is entered as distance to the end point. Within the dynamics
iodine will be produced with

Min. rate

. The bigger the control range the

slower the titration. Outside the dynamics iodine will be produced continu-
ously with

Max. rate

.

The standard control parameters are optimal for most applications and
should not be altered. If you nevertheless need to alter the control parame-
ters for special reagents and/or samples take care that the polarization cur-
rent of the indicator electrode (

I(pol)

), the end point and the dynamics are

linked to each other.