Metrohm tiamo 1.0 Manual User Manual
Page 426

402
Method
Electrode test
on, [ off ]
Switching on/off the electrode test for polarized electrodes. The electrode
test for polarized electrodes is carried out during the transition from an in-
active condition to a measuring condition
Generator electrode
Generator type
[ without diaphragm ], with diaphragm
Selection of the generator type.
Generator current
100, 200, [ 400 ], mA, auto
Polarization current at the generator electrode.
auto
means that the current
is automatically adapted to the conductivity of the reagent and that in the
region of the endpoint the current will be controlled at smaller values.
Note
For generator electrodes
with diaphragm
it is recommended to set the
Generator
current
to
auto
.
For generator electrodes
without diaphragm
it is recommended to set the
Gene-
rator current
to
400 mA
.
Stirrer
Switch on/off automatically
[ on ], off
If this option is eanbled, the stirrer will be switched on automatically at the
start of the command and will be switched off automatically at the end of
the command.
Start conditions
Command name
25 characters, [ KFC # ]
Name of the command.
Pause
Pause
[ 0 ] ... 999999 s
Waiting time before the start of the titration.