beautypg.com

Metrohm tiamo 1.0 Manual User Manual

Page 44

background image

20

General program functions

Identifier

Description

Commands

.EP{#}.MEP

Number of end points in the window # (1...9); 1 = 1 end point,

2 = 2 or more end points

DET, MET, SET, KFT

.EP{#}.QTY

Measured value (water) for end point # (1...9) in µg

KFC

.EP{#}.TEM

Temperature for end point # (1...9) in °C

DET, MET. SET, KFT

.EP{#}.TIM

Time for reaching end point # (1...9) in s

DET, MET, SET,

KFT, KFC

.EP{#}.VOL

Volume for end point # (1...9) in mL

DET, MET, SET, KFT

.ETE

Final temperature (temperature after processing the command)
in °C

DET, MET, SET,
KFT, KFC, MEAS
(without T and

T/FLOW), CAL
MEAS

.EVT

Final volume (total volume added at the end of the command) in
mL

DET, MET, SET, KFT

.FIN

Command status; 1 = command has been finished at least
once
, 0, invalid (variable not present) = command has never

been finished

All

.FP{#}.DME

Difference in measured values for fixed end point # (1...9)

MET

.FP{#}.MEA

Measured value for fixed end point # (1...9) in mV

KFC, MEAS (without
T/Flow)

.FP{#}.MEA

Measured value for fixed end point # (1...9) in the units of the

measured value

DET, MET, SET, KFT

.FP{#}.DVT

Drift for fixed end point # (1...9) in µg/min

KFC

.FP{#}.ERC

ERC for fixed end point # (1...9)

DET

.FP{#}.QTY

Measured value (water) for fixed end point # (1...9) in µg

KFC

.FP{#}.TEM

Temperature for fixed end point # (1...9) in °C

DET, MET, SET,

KFT, MEAS (without
T and T/Flow)

.FP{#}.TIM

Time for reaching fixed end point # (1...9) in s

DET, MET, SET,
KFT, KFC, MEAS
(without T/Flow)

.FP{#}.VOL

Volume for fixed end point # (1...9) in mL

DET, MET, SET, KFT

.HP{#}.MEA

Measured value for HNP # (1...9) in mV (HNP = half-

neutralization potential)

DET, MET

.HP{#}.TEM

Temperature for HNP # (1...9) in °C

DET, MET

.HP{#}.TIM

Time for reaching HNP # (1...9) in s

DET, MET

.HP{#}.VOL

Volume for HNP # (1...9) in mL

DET, MET

.IGF

Initial gas flow (measured value at start of command) in mL/min

MEAS T/Flow

.IME

Initial measured value (measured value before processing the
start condition) in the units of the measured value

DET, MET, SET,
KFT, MEAS, CAL
MEAS

.ITE

Initial temperature (temperature before processing the start con-

ditions) in °C

DET, MET, SET,

KFT, MEAS (without
T and T/Flow), CAL
MEAS