beautypg.com

Metrohm tiamo 1.0 Manual User Manual

Page 288

background image

264

Database

y1-axis

Command type

[ DET ], MET, SET, KFT, KFC, MEAS, MEAS T/Flow

Selects the command type for which the curve properties are to be defined.
The curve properties defined per command type are saved per curve win-
dow and per client.

Note

When the dialog window is opened the particular

Command type

that will be se-

lected as standard is that which applies for the curve shown in the curve win-
dow.

Autoscaling

[ on ], off

If this option is switched on then all axes in the curve window will be scaled

automatically. In this case the fields

Start value

and

End value

cannot be ed-

ited.

Parameters for the graphical presentation of the curves on the y1 axis (left-hand y
axis).

Quantity

Command-dependent selection

Selection of the quantity to be shown on the y1 axis.

Label

50 characters, [ auto ]

Freely definable axis labeling for the y1 axis. With

auto

the label from the

field

Quantity

will be used.

Scaling

Start value

1 E-8 ... [ 0.0 ] ... 1 E+8

Initial value for scaling the y1 axis.

End value

1 E-8 ... [ 1000.0 ] ... 1 E+8

Final value for scaling the y1 axis.