beautypg.com

Metrohm tiamo 1.0 Manual User Manual

Page 137

background image

Subwindow Run

113

Cancel sequence

A running series can be canceled at any time with

[Stop]

. An Exit track

may be carried out; the determination will be terminated.

Cancel by error

If when a determination is being carried out an error occurs that
causes the cancellation of the determination then an Error track may
be carried out; the determination will be terminated.

Process exit track

When the main track and all the other tracks that it has called up have
been completed then an Exit track may be carried out; the determina-
tion will be terminated. After the end of the Main track the status
changes to

READY

.

7 - Post-conditioning (only for methods with conditioning)

In methods that contain commands with switched-on conditioning and in which
the option

Automatic conditioning

is switched on in the

START

command then for all

these commands post-conditioning will be started automatically with the loaded
method immediately after the end of the track that contains the conditioning
command.

Conditioning requirement not fulfilled

After the start of post-conditioning the status changes to

COND BUSY

.

In this status a new determination can be started with

[Start]

, even

though the conditioning requirement has not been fulfilled. Condition-
ing can also be canceled with

[Stop]

or interrupted with

[Hold]

and

then continued with

[Cont]

. If an error occurs during conditioning then

neither the Error track nor the Exit track will be carried out.

Conditioning requirement fulfilled

If the conditioning requirement has been fulfilled then the status
changes to

COND READY

. In this status a new determination can be

started with

[Start]

. Conditioning can also be canceled with

[Stop]

or

interrupted with

[Hold]

and then continued with

[Cont]

. The condition-

ing requirement will continue to be monitored continuously until the
determination has been started.

Device assignment

If at the Start test it is discovered that no device has been assigned for a device-
specific command then the window

Device assignment

appears for assigning the

device:

Command type

Shows the command type.

Command name

Shows the command name.

Device type

Shows the device type selected in the command.

Device name

Device name

Selection of a device from the device table. Only those devices being able
to execute the command are selectable.