Determination run – Metrohm tiamo 1.0 Manual User Manual
Page 135

Subwindow Run
111
Reason
Selection from standard texts
Slection from the Default reasons defined in the dialog window
Security set-
tings
for the category
Modification of sample data
.
Comment
1000 characters
Entry of comments about the modification to the sample data.
Determination run
A determination that is started on the tab
Single determination
in the subwindow
Run
takes place as follows:
1 - Load sample data
The Sample data entered on the tab
Single determination
is loaded for the deter-
mination.
2 - Select method
Method selection triggers the following actions:
• Select method without assignment ID
If the option
Method selection by assignment ID
is not switched on on
the tab
Process
then the method selected in the field
Method
will remain
loaded.
• Select method with assignment-ID
If the option
Method selection by assignment ID
is switched on on the
tab
Process
then the
Sample assignment ID
will be used to load the
method defined in the
Sample assignment table
.
3 - Carry out start test
During the Start test the following checks and actions are triggered:
• Right to carry out the method?
Checks whether the logged-in user has the right to execute the par-
ticular method.
• Method executable?
Checks whether the particular method is executable. If the method is
non-executable then a message appears asking for the Method test to
be carried out and the method edited accordingly.
• Check device (module) and rack
Checks whether the devices (or device types), modules (e.g. measur-
ing inputs, towers, etc.) and racks demanded in the device-specific
commands are present and available or assigned. If the device has
not yet been assigned in a command then the window
Device assign-
ment
appears for assigning the device.
• Check solutions
Checks whether the solutions demanded in the device-specific com-
mands are present and available.
• Check sensors
Checks whether the sensors demanded in the device-specific com-
mands are present and available.
Note
The Start test can also be triggered manually in the
READY
status with the menu
item
Tools, Run test
.