Determination run – Metrohm tiamo 1.0 Manual User Manual

Page 135

background image

Subwindow Run

111

Reason

Selection from standard texts

Slection from the Default reasons defined in the dialog window

Security set-

tings

for the category

Modification of sample data

.

Comment

1000 characters

Entry of comments about the modification to the sample data.

Determination run

A determination that is started on the tab

Single determination

in the subwindow

Run

takes place as follows:

1 - Load sample data

The Sample data entered on the tab

Single determination

is loaded for the deter-

mination.

2 - Select method

Method selection triggers the following actions:

Select method without assignment ID

If the option

Method selection by assignment ID

is not switched on on

the tab

Process

then the method selected in the field

Method

will remain

loaded.

Select method with assignment-ID

If the option

Method selection by assignment ID

is switched on on the

tab

Process

then the

Sample assignment ID

will be used to load the

method defined in the

Sample assignment table

.

3 - Carry out start test

During the Start test the following checks and actions are triggered:

Right to carry out the method?

Checks whether the logged-in user has the right to execute the par-
ticular method.

Method executable?

Checks whether the particular method is executable. If the method is

non-executable then a message appears asking for the Method test to
be carried out and the method edited accordingly.

Check device (module) and rack

Checks whether the devices (or device types), modules (e.g. measur-
ing inputs, towers, etc.) and racks demanded in the device-specific

commands are present and available or assigned. If the device has
not yet been assigned in a command then the window

Device assign-

ment

appears for assigning the device.

Check solutions

Checks whether the solutions demanded in the device-specific com-

mands are present and available.

Check sensors

Checks whether the sensors demanded in the device-specific com-
mands are present and available.

Note

The Start test can also be triggered manually in the

READY

status with the menu

item

Tools, Run test

.