Commands
443
Temperature
Reference temperature
-170.0 ... [ 20.0 ] ... 500.0 °C
The electrical conductivity depends greatly on the temperature. The con-
ductivity measured at a particular temperature is thus automatically con-
verted to this reference temperature (usually 20 °C or 25 °C).
Measurement temperature
-170.0 ... [ 20.0 ] ... 500.0 °C
Temperature of the solution to be measured. If a temperature sensor is
connected, the temperature entered here is ignored.
Evaluations
Command name
25 characters, [ MEAS Cond # ]
Name of the command.
The following methods for evaluating measuring curves are defined on this tab.
Fix end point evaluation
on, [ off ]
To a fixed quantity (measured value or time) the associated values for the
other quantities are interpolated from the list of measuring points for the
fixed end point.
Fixed quantity
[ Measured value ], Time
Selection of the fixed quantity to which the associated values for the other
quantities are to be interpolated from the list of measuring points.
Fix EP# at
0.0000 ... 2'000.0 mS/cm, [ off ]
Fixed end point 1 ... 9 for
Fixed quantity
=
Measured value
.
Fix EP# at
0.0 ... 999'999.9 s, [ off ]
Fixed end point 1 ... 9 for
Fixed quantity
=
Time
.