Method

SET

Command for Set Endpoint Titrations (

SET

).

Principle

With this command titrations to one or two preset endpoints are carried out. The
termination of the titration at the endpoint is either drift-controlled or after a waiting
period. The volume that has been added when the endpoint is reached gives the

calculable reagent consumption.

Reagent addition takes place in three phases during the titration:

Initial addition

During this phase the dosing rate increases continuously. It starts with
the

Minimum rate

and increases to the

Maximum rate

.

Continuous addition

In this phase addition is carried out at the

Maximum rate

until the

Con-

trol range

is reached.

Control range

In this range addition is finely controlled. Shortly before the end point
is reached addition is only carried out at the

Minimum rate

(see "Con-

trol range").

Note

This command is used for rapid routine determinations when the end point does
not change throughout a series or when an excess of reagent must be avoided.

Commands

Depending on the measured value one of the following four

SET

commands can

be chosen: