Metrohm tiamo 1.0 Manual User Manual
Page 456

432
Method
Device
Device name
Device name, [ not defined ]
Selection of a device from the device table. Only those devices being able
to execute the command are selectable. Additionally,
not defined
can be se-
lected. In this case, the device name must be assigned by the user at the
start of the method.
Device type
Device types, [ Titrando ]
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.
If the option
not defined
is selected as
Device name
, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Sensor
Measuring input
[ 1 ] ... 2
(Titrando)
[ 1 ] ... 2, diff.
(Titrino)
Selection of the measuring input to which the sensor is connected.
Sensor
Sensor name, [ Temperature sensor ], pH electrode
Selection of a pH sensor from the pH sensors configured in the sensor ta-
ble.
Stirrer
Stirrer
[ 1 ] ... 4, off
(Titrando only)
Selection of the stirrer.
off
means that no stirrer will be used.
Stirring rate
-15 ... [ 8 ] ... 15
(Titrando only)
Setting the stirrer speed. The direction in which stirring takes place changes
as the sign in front of the stirrer speed changes.
Switch off automatically
[ on ], off
(Titrando only)
The stirrer is automatically switched off at the end of the command if this
option is activated.
Measuring parameters