Metrohm tiamo 1.0 Manual User Manual

Page 456

background image

432

Method

Device

Device name

Device name, [ not defined ]

Selection of a device from the device table. Only those devices being able

to execute the command are selectable. Additionally,

not defined

can be se-

lected. In this case, the device name must be assigned by the user at the
start of the method.

Device type

Device types, [ Titrando ]

Display or selection of the device type. If a device is selected under

Device

name

, this parameter is not editable any more. The device type belonging to

the selected device is displayed instead.
If the option

not defined

is selected as

Device name

, all device types being

able to execute the command can be selected independently from the de-
vices in the device table.

Sensor

Measuring input

[ 1 ] ... 2

(Titrando)

[ 1 ] ... 2, diff.

(Titrino)

Selection of the measuring input to which the sensor is connected.

Sensor

Sensor name, [ Temperature sensor ], pH electrode

Selection of a pH sensor from the pH sensors configured in the sensor ta-
ble.

Stirrer

Stirrer

[ 1 ] ... 4, off

(Titrando only)

Selection of the stirrer.

off

means that no stirrer will be used.

Stirring rate

-15 ... [ 8 ] ... 15

(Titrando only)

Setting the stirrer speed. The direction in which stirring takes place changes
as the sign in front of the stirrer speed changes.

Switch off automatically

[ on ], off

(Titrando only)

The stirrer is automatically switched off at the end of the command if this

option is activated.

Measuring parameters