Page 413

background image

Commands

389

If the EP has been reached, after the last dosing the delay time here de-
fined is awaited and then the titration is stopped. This parameter is only

shown for

Stop criterion

=

time

.

inf.

stands for infinite.

Stop time

1 ... 999999 s

(only Titrino 720, 736, 751, 758, 784, 795, 799)

The titration is stopped if the defined stop time has elapsed. This parameter
is only shown for

Stop criterion

=

time

and

Delay time

=

inf.

Endpoint 2

(only visible for

EP1 bei

off

)

[ on ], off

Switching endpoint 2 on/off. If this is deactivated all following parameters
are not visible.

EP2 at

-200.0 ... 200.0 µA, [ off ]

Measured value for the second endpoint.
Parameters for

Endpoint 2

see above.

Titration parameters

Screenshot and parameters: see SET pH - Titration parameters

Stop conditions

Screenshot and parameters: see SET pH - Stop conditions

Conditioning

Screenshot and parameters: see SET pH - Conditioning

Additional evaluations