Metrohm tiamo 1.0 Manual User Manual
Page 601

Subwindow Devices
577
Request for dosing device preparation
Selection when the request for carrying out a
PREP
(prepare) command is to be
shown for the internal dosing device.
At program start
[ on ], off
If this option is switched on then the request to prepare the dosing device
will appear at each program start.
On attaching an exchange/dosing unit
[ on ], off
If this option is switched on then the request to prepare the dosing device
will appear each time that the exchange unit is attached to the Titrino.
Time interval
on, [ off ]
If this option is switched on then the request to prepare the dosing device
will appear after the time interval defined here.
all
0.1 ... [ 12 ] ... 999.9 h
Time interval after which the request to prepare the dosing device will ap-
pear.
Ext. dosing device D#
Request for dosing device preparation
Selects when the request for carrying out a
PREP
(prepare) command is to be
shown for the external dosing device
D1
or
D2
.
At program start
[ on ], off
If this option is switched on then at each program start the request to pre-
pare the dosing device will appear.
On attaching an exchange/dosing unit
[ on ], off
If this option is switched on then each time that an exchange unit is at-
tached to the Dosimat or each time that a dosing unit is attached to the
Dosino the request to prepare the dosing device will appear.
Time interval
on, [ off ]
If this option is switched on then the request to prepare the dosing device
will appear after the time interval defined here.
all
0.1 ... [ 12.0 ] ... 999.9 h
Time interval after which the request to prepare the dosing device will ap-
pear.