Commands

371

Max. waiting time

0 ... [ 1 ] ... 999999 s

(only Titrando)

If signal drift has been switched off or has not yet been achieved then the
measured value will be accepted when the maximum waiting time has
elapsed.

Pause 1

Pause 1

[ 0 ] ... 999999 s

(only Titrando and Titrino 719, 720, 736, 751, 758, 785, 794,

798, 799)

Waiting time, e.g. until the electrode has settled down, before a start vol-
ume is added.

Start volume

Start volume

[ 0.00000 ] ... 9999.99 mL

(Titrando),

[ 0.00 ] ... 999.99 mL

(Titrino)

Volume to be added before the start of the titration with the below entered
speed.

Dosing rate

0.01 ... 166.00 mL/min, [ maximum ]

(Titrando),

0.01 ... 150.00 mL/min, [ maximum ]

(Titrino)

Speed at which the start volume is to be added. The maximum dosing rate
depends on the cylinder volume of the used exchange unit or dosing unit.

Pause 2

Pause 2

[ 0 ] ... 999999 s

Waiting time, e.g. until the electrode has settled down after the start or a re-
action time after the addition of a start volume. The pause takes place at

the end of all start conditions.