Metrohm tiamo 1.0 Manual User Manual

Page 450

background image

426

Method

Command name

25 characters, [ MEAS Upol # ]

Name of the command.

The general parameters for the control device, the sensor and the stirrer are de-
fined on this tab.

Device

Device name

Device name, [ not defined ]

Selection of a device from the device table. Only those devices being able
to execute the command are selectable. Additionally,

not defined

can be se-

lected. In this case, the device name must be assigned by the user at the
start of the method.

Device type

Device types, [ Titrando ]

Display or selection of the device type. If a device is selected under

Device

name

, this parameter is not editable any more. The device type belonging to

the selected device is displayed instead.
If the option

not defined

is selected as

Device name

, all device types being

able to execute the command can be selected independently from the de-
vices in the device table.

Sensor

Measuring input

[ 1 ] ... 2

(Titrando)

[ 1 ] ... 2, diff.

(Titrino)

Selection of the measuring input to which the sensor is connected.

Sensor

Sensor name, [ Metal electrode ], Conductivity sensor

Selection of a sensor from the sensors configured in the sensor table.

U(pol)

-1'250 ... [ 400 ] ... 1'250 mV

(Titrando, increment = 25 mV)

-1'270 ... [ 400 ] ... 1'270 mV

(Titrino, increment = 10 mV)

The polarization voltage is the potential applied to the polarized electrode
during a amperometric measurement.

Electrode test

on, [ off ]

Switch on/off the electrode test for polarized electrodes. This test is carried
out during the transition from an inactive condition to a measuring condi-
tion. It is checked that the electrode is properly connected and that no
short-circuit is present.

Temperature measurement

Type of temperature measurement (Titrando only):

continuous

A temperature sensor must be connected. The temperature is then meas-
ured continuously.

[ automatic ]

If a temperature sensor is connected, the temperature will be measured
continuously. Otherwise the temperature manually entered under

Tempera-

ture

(see Measuring parameters) will be used.

off

The temperature will not be measured. The temperature manually entered
under

Temperature

(see Measuring parameters) will be used.