374

Method

EP2 at pH

-20.000 ... 20.000, [ off ]

(Titrando),

-20.00 ... 20.00, [ off ]

(Titrino)

Measured value for the second endpoint.
Parameters for endpoint 2 see above.

Titration parameters

Command name

25 characters, [ SET pH # ]

Name of the command.

Parameters defining the run of the titration.

Titration direction

Selection of the titration direction. If two endpoints are set the direction is al-

ready given and the following selection is ignored.

+

positive measured value alteration

[ - ]

negative measured value alteration

auto

The titration direction is determined automatically from the start measured

value and the set endpoint.

Extraction time

0 ... 999999 s

(only Titrando and Titrino 719, 720, 736, 751, 758, 785, 794,

798, 799)
The titration is not stopped until the extraction time has elapsed (even when
the EP has already been reached). For example, the entry of an extraction

time is a good idea for the titration of sparingly soluble samples.

Temperature

-20.0 ... [ 25.0 ] ... 150.0 °C

(Titrando)

-170.0 ... [ 25.0 ] ... 500.0 °C

(Titrino)

Manually entered titration temperature. If a temperature sensor is con-

nected and the

Temperature measurement

in the tab

General/Hardware

under

Sensor

is set to

automatic

or

continuous

then the temperature will be meas-

ured continuously.

Time interval measuring point

0.1 ... [ 2.0 ] ... 999999 s

(Titrando)

0.08 ... [ 2.00 ] ... 16200 s

(Titrino)

Time interval for entering a measuring point in the list of measuring points.