Upper limit
0.0000 ... [ 2'000.0 ] mS/cm
(for
Window
=
Measured value
)
Measured value for the upper limit of the window.
0 ... [ 999'999] s
(for
Window
=
Time
)
Time for the upper limit of the window.
Additional measured values
Screenshot and parameters: see DET pH - Additional measured values
5.6.7 Calibration
commands
Commands for calibration of pH- and ion selective sensors.
General
The commands
CAL LOOP
and
CAL MEAS
are available for calibrating electrodes.
The number and type of buffers or standards respectively is defined in the loop
command
CAL LOOP
whereas the measurement is done with the command
CAL
MEAS
. The validity of the measurement is checked after each pass of the loop. The
calibration data for the electrode are determined at the end.
The basic structure of a calibration method looks as follows:
The calibration can be carried out by manual changing of buffers/standards or by
automatic changing in combination with a Sample Processor.
Commands
The following commands can be selected for calibration:
•
CAL LOOP pH
Calibration loop for calibrating pH electrodes.
•
CAL MEAS pH
Measuring command for calibrating pH electrodes.