Metrohm tiamo 1.0 Manual User Manual

Page 482

background image

458

Method

Stirrer

Stirrer

[ 1 ] ... 4, off

(only Titrando, 846, 814, 815)

Selection of the stirrer.

off

means that no stirrer will be used.

Stirring rate

-15 ... [ 8 ] ... 15

(only

Titrando, 846, 814, 815)

Setting the stirrer speed. The direction in which stirring takes place changes
as the sign in front of the stirrer speed changes.

Switch off automatically

[ on ], off

(only

Titrando, 846, 814, 815)

The stirrer is automatically switched off at the end of the command if this
option is enabled.

Parameters

Command name

25 characters, [ Dosing # ]

Name of the command.

The volume to be added and the dosing and filling rate are defined on this tab.

Volume

0.00000 ... [ 10.0000 ] ... 99999.9 mL

Volume to be added.

Dosing rate

[ maximum ]
0.01 ... 166.00 mL/min

(Titrando, 814 815)

0.01 ... 160.00 mL/min

(730, 774, 778, 789)

0.01 ... 150.00 mL/min

(Titrino)

Speed at which dosing is carried out. The maximum dosing rate depends
on the cylinder volume of the exchange unit or dosing unit used. In case the
entered dosing rate is too high for the selected dosing device, the rate will

be reduced automatically to the highest possible value.

Note
Lower dosing rates should be entered for solutions with higher viscosities.

Filling rate

[ maximum ]
0.01 ... 166.00 mL/min

(Titrando, 814 815)

0.01 ... 160.00 mL/min

(730, 774, 778, 789)

0.01 ... 150.00 mL/min

(Titrino)

Speed at which the buret is to be filled after dosing. The maximum filling
rate depends on the cylinder volume of the exchange unit or dosing unit
used. In case the entered filling rate is too high for the selected dosing de-

vice, the rate will be reduced automatically to the highest possible value.