Send – Metrohm tiamo 1.0 Manual User Manual
Page 518

494
Method
If the option
not defined
is selected as
Device name
, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Remote box
[ 1 ] ... 4
(only Titrando, 814, 815)
Selection of the remote box at which the lines are to be scanned.
Scan lines
Input signal
Binary pattern of exactly 8 characters (0, 1,*), [ ******** ], signal template
Input of the binary pattern for the expected input signal or selecting a pre-
defined Signal pattern.
The following characters can be entered:
0
= Line inactive
1
= Line active
*
= Any line condition
The bits are always numbered from right to left:
Input 7 6 5 4 3 2 1 0
Bit
7 6 5 4 3 2 1 0
Example:
*******1
is waiting for an active input on line 0 (1 = set). For
example, this line is set by a Titrino after a titration has been completed and
the Titrino can accept a start signal again.
Note
Input lines that are of no interest or for which no defined condition can be pre-
dicted should also be masked with an asterisk (
*
).
Timeout
on, [ off ]
Maximum waiting time for the input signal.
SEND
Command for sending event messages to commands, tracks or to tiamo.
Appearance
The command has the following appearance:
Parameters
The parameters for the command
SEND
are configured in the following dialog win-
dow:
•
SEND