beautypg.com

Metrohm tiamo 1.0 Manual User Manual

Page 666

background image

642

Configuration

Dosing port 1
Port

[ Port 1 ], Port 2, Port 3, Port 4

Port to be used as Dosing port 1.

Length

0.0 ... [ 40.0 ] ... 999.9 cm

Length of tubing attached to Dosing port 1.

Diameter

0.0 ... [ 2.0 ] ... 9.9 mm

Diameter of tubing attached to Dosing port 1.

Dosing port 2
Port

Port 1, Port 2, [ Port 3 ], Port 4

Port to be used as Dosing port 2.

Length

[ 0.0 ] ... 999.9 cm

Length of tubing attached to Dosing port 2.

Diameter

0.0...[ 2.0 ]...9.9 mm

Diameter of tubing attached to Dosing port 2.

Fill port
Port

Port 1, [ Port 2 ], Port 3, Port 4

Port to be used as Fill port for aspirating the solution.

Length

0.0...[ 25.0 ]...999.9

Length of tubing attached to Fill port.

Diameter

0.0...[ 2.0 ]...9.9 mm

Diameter of tubing attached to Fill port.

Special port

Port

Port 1, Port 2, Port 3, [ Port 4 ]

Port to be used as Special port.

Length

[ 0.0 ] ... 999.9 cm

Length of tubing attached to Special port.

Diameter

0.0...[ 2.0 ]...9.9 mm

Diameter of tubing attached to Special port.

Port occupancy of dosing unit: