Metrohm tiamo 1.0 Manual User Manual
Page 686

662
Configuration
With variables of the type
Date/Time
the date can be selected by pressing
on
in the dialog window
Select date
(see General program functions -
Edit - Select date).
Value (unit)
20 characters, 'empty', [ mol/L ], mmol/L, µmol/mL, g/L, mg/L, µg/mL, mg/mL,
ppm, %, mEq/L
Name of the unit. If a value is assigned automatically by a
CALC
command
then the unit will be automatically entered here. It is available for methods
that use the common variable as variable
'CV.Name.UNI'
for calculations.
Comment
256 characters
It is possible to enter remarks about the common variable.
Assignment date
Date and time at which the last value was assigned; this is entered auto-
matically each time that a value is assigned automatically or manually.
Note
When assigning the value manually the date is only entered if the value has really
been changed.
Assignment method
Name of the method with which the last value assignment was carried out.
If the value was entered manually then
manual
will be shown here.
User
Short name of the user who was logged in during value assignment or who
entered the value manually. If work is not carried out under login then the
name of the user logged in under Windows will automatically be entered
here.
Common variable monitoring
on, [ off ]
If this option is switched on then the validity of the common variables will be
monitored.
Validity
0 ... [ 999 ] days
Period that the common variable is valid in days. If a value is entered here
then the field
Next assignment
will be adapted automatically.
Next assignment
Date selection, [ assignment date + 999 days ]
Date on which the next value assignment must take place. This date can be
selected by pressing on
in the dialog window
Select date
(see General
program functions - Edit - Select date). When the date has been entered the
value for the
validity
period will be adapted automatically.
Action
If during common variable monitoring it is found that the validity period has ex-
pired then one of the following actions will be triggered automatically at Start test:
Record message
The message that the validity period of the common variable has expired
will be automatically saved with the determination.
Display message
A message will be shown and you can then choose if you wish to continue
with the run or terminate it. If the run is continued then the message that the
validity period of the common variable has expired will be automatically
saved with the determination.