442

Method

Device

Device name

Device name, [ not defined ]

Selection of a 712 Conductometer from the device table. Additionally,

not

defined

can be selected. In this case, the device name must be assigned by

the user at the start of the method.

Device type

712 Conductometer

Display or selection of the device type. If a device is selected under

Device

name

, this parameter is not editable any more. The device type belonging to

the selected device is displayed instead.
If the option

not defined

is selected as

Device name

, all device types being

able to execute the command can be selected independently from the de-
vices in the device table.

Sensor

Sensor

Sensor name, [ Conductivity sensor ]

Selection of a conductivity sensor from the conductivity sensors configured
in the sensor table.

Measuring parameters

Command name

25 characters, [ MEAS Cond # ]

Name of the command.

The measuring parameters are defined on this tab.

Measurement

Measurement frequency

[ auto ], 300 Hz, 2.4 kHz

The more suitable frequency is chosen with the setting

auto

.

Measuring time

0 ... [ 120 ] ... 999'999 s

Maximum period of time for measurement.

Measuring interval

1 ... [ 2 ] ... 999'999.0 s

Time interval for entering a measuring point in the list of measuring points.

Stop measured value

0.0000 ... 2000.0 mS/cm, [ off ]

Stop when the preset measured value has been reached since the start of
the measurement.

off

means no stop.