Commands

447

Calibrating with automatic solution changing

The basic structure of a calibration method with automatic solution exchange on a
Sample Processor looks as follows:



Defining the number of buff-
ers/standards and type of solution
exchange.


Move to the next buffer/standard.

The request for changing the buffer/standard must be deactivated in the com-
mand

CAL LOOP pH

or

CAL LOOP Conc

respectively. No request for changing the

buffer/standard is displayed during the method run.

For the automatic move to the next calibration solution on the Sample Processor a
command

MOVE

must be inserted between the loop command

CAL LOOP

and the

measuring command

CAL MEAS

. The positions of the calibration solutions on the

sample rack are defined in the command

MOVE

. There are three possibilities:

Defining rack positions for calibration solutions in the method

A

Rack position

###

for the first calibration solution has to be defined

in the command

MOVE

. The other calibration solutions must be placed

on the directly following rack positions. The following settings have to
be done in the command

MOVE

under the parameter

Target

:

Move

:

Rack position

Number

:

= ### – 1 + 'Command name.LCO'

(entry with Formula editor,

i.e.

= 19 + 'Calibration loop pH.LCO'

for the above sample method with

the first calibration solution on rack position

20

).

Defining special beakers for calibration solutions in the method

A

Special beaker

###

for the first calibration solution has to be defined

in the command

MOVE

. The other calibration solutions must be placed

on the following special beaker positions. As many special beakers
must be defined for the sample rack as buffers or standard solutions
are required for the calibration (see Rack properties - Special beak-
ers). The following settings have to be done in the command

MOVE