498

Method

Open the dialog window

Wait for event/status

for entering a new event

message being waited for.

Note
10 event messages can be entered at maximum.

Open the dialog window

Wait for event/status

to edit the selected event

message.

Deleting the selected event message.

Timeout

[ on ], off

[ 0 ] ... 9999.9; [ min ], s

Optional entry of a maximum waiting time. In case none of the conditions

defined above are fulfilled after this time period, the command is cancelled
and the proximate command is started.

Message

This message is displayed in the live window while waiting for the condition
defined above. The Text editor to enter or change the message is opened

with

or by double-clicking into the message field.

SEND - event/state

Command

Command

Name of the command whose event message is waited for.

Event message

Event message, [ 'empty' ]

Selecting the event message/status to be waited for. The following event

messages and status can be selected:

Event/Status Type

Meaning

Commands

Start

Event Command has been started.

All

End

Event Command has been finished.

All

Dosing

Event The dosing of the titration has been started.

DET

,

MET

Measure

Event The measurement of the titration has been started.

DET

,

MET

Busy

Status System is in condition

BUSY

,

HOLD

or

ERROR

. All

Finished

Status Command has been finished at least once.

All