Metrohm tiamo 1.0 Manual User Manual

Page 638

background image

614

Configuration

Dosing device

Shows the number and type of dosing devices connected to the 730 Sample
Changer
or 774 Oven Sample Processor. A maximum of 12 type

685 Dosimat

or

700 Dosino

dosing devices can be connected.

Note
Connecting up the dosing devices is described in the Instructions for Use for the
730 Sample Changer and 774 Oven Sample Processor. If no dosing device is
connected then the card will remain empty.

Oven

On the tab

Oven

the oven parameters for the 774 Oven Sample Processor are

shown. The parameters cannot be edited, they are only read off by the device.

Initial temperature

Shows the initial temperature set on the device. Entering an initial tempera-
ture means that the oven will heat up to the set temperature when it is
switched on.

off

means that the oven will not be switched on.

Max. temperature

Shows the maximum temperature set on the device. This parameter is a
safety setting and is intended to prevent a sample from being overheated. If
the oven reaches this temperature on heating up then the oven heating will
be switched off.