Commands
381
Stop drift
1 ... [ 20 ] ... 999 µL/min
If the EP and the stop drift have been reached the titration will be stopped.
This parameter is only shown for
Stop criterion
=
drift
.
Delay time
0 ... [ 10 ] ... 999 s
(Titrando 835, 836, 841),
0 ... [ 10 ] ... 999 s, inf.
(Titrino 720, 736, 751, 758, 784, 795, 799)
If the EP has been reached, after the last dosing the delay time here de-
fined is awaited and then the titration is stopped. This parameter is only
shown for
Stop criterion
=
time
.
inf.
stands for infinite.
Stop time
1 ... 999999 s
(only Titrino 720, 736, 751, 758, 784, 795, 799)
The titration is stopped if the defined stop time has elapsed. This parameter
is only shown for
Stop criterion
=
time
and
Delay time
=
inf.
Endpoint 2
(only visible for
EP1 bei
≠
off
)
[ on ], off
Switching endpoint 2 on/off. If this is deactivated all following parameters
are not visible.
EP2 at
-2000.0 ... 2000.0 mV, [ off ]
(Titrando),
-2000 ... 2000, [ off ]
(Titrino)
Measured value for the second endpoint.
Parameters for endpoint 2 see above.
Titration parameters
Screenshot and parameters: see SET pH - Titration parameters
Stop conditions
Screenshot and parameters: see SET pH - Stop conditions
Conditioning
Screenshot and parameters: see SET pH - Conditioning