SET pH
Potentiometric pH measurement with pH electrodes (measured quan-
tity pH).
•
SET U
Potentiometric voltage measurement with metal electrodes (measured
quantity voltage U).
•
SET Ipol
Voltametric measurement with selectable polarization current (meas-
ured quantity voltage U).
•
SET Upol
Amperometric measurement with selectable polarization voltage
(measured quantity current I).
SET pH
Command for Set Endpoint Titrations with potentiometric pH measurement.
Devices
This command can be executed with the following devices which have the SET
mode:
Titrino: 702, 716, 718, 719, 720, 721, 736, 751, 758, 785, 794, 798, 799
Titrando: 808, 809, 835, 836, 841
Appearance
The command has the following appearance:
without conditioning with conditioning
Parameters
The parameters for the command
SET pH
are configured in the following 8 tabs:
•
General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.
•
Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.
•
Control parameters
Control parameter for endpoint 1 and endpoint 2 (measured value for
the endpoint, titration rate, user defined parameters, stop criterion).
•
Titration parameters
Parameters for the run of the titration.
•
Stop conditions
Definitions of conditions which cause the stop of the titration.
•
Conditioning
Switching conditioning on/off. Definition of start drift, drift correction
and stop conditions.
•
Additional evaluations
Definition of additional methods for evaluation of the titrations curves.