Page 491

background image

Commands

467

Relative angle

(only for 778, 789, 814, 815)

The sample rack can be rotated by a certain amount relatively to the current

position independently from rack positions. This can be used for example
to remove vial caps automatically.

Number

[ 1 ] ... 999

Specification of the number of the rack position. This parameter is only visi-
ble for

Move

=

Rack position

or

Special beaker

. The rack position can be en-

tered as a formula as well. This is necessary for example for the automatic
calibration of electrodes. The Formula editor (see Formula editor) is opened
by right clicking into the input field.

Move angle

-259.90 ... [ 5.00 ] ... 259.90 °

This parameter specifies the angle about which the sample rack should be

rotated relatively to the current position. It is only visible for

Move

=

Relative

angle

.

Beaker test

Selection of the action to be carried out if the

Beaker sensor

defined for the

rack doesn't detect a sample at the target position. If the

Robotic arm

is de-

fined as beaker sensor, a Swing Head must be installed and a suitable

Work position

with beaker contact must be defined for the lift. This position

will be moved to automatically for beaker recognition after the command

MOVE

. This section is not visible for

Move

=

Relative angle

.

Display message

[ on ], off

A message is displayed and the determination is stopped.

Stop determination

on, [ off ]

The determination is stopped and the next determination of the series is
started.

Stop determination and series

on, [ off ]

The determination as well as the series are stopped. An Error track will be

started if there is one defined.

Parameters

Shift rate

3 ... [ 20 ] °/s

Entry of the speed at which the sample rack rotates.

Shift direction

[ auto ], +,

Selection of the direction of rotation of the sample rack. The direction of ro-

tation with the smallest angle of rotation is selected automatically with

auto

.

The direction of rotation is counterclockwise with "

+

" and clockwise with "

-

".

Swing rate

10 ... [ 55 ] °/s

Entry of the speed of the robotic arm when moving to a rack position or a
special beaker (only with multiple row sample racks).