Metrohm tiamo 1.0 Manual User Manual
Page 458

434
Method
Measuring time
0 ... [ 120 ] ... 999'999 s
Maximum period of time for measurement.
Measuring interval
0.1 ... [ 2.0 ] ... 999'999.0 s
(Titrando, increment = 0.1 s)
0.08 ... [ 2.0 ] ... 16'200 s
(Titrino, increment = 0.08 s)
Time interval for entering a measuring point in the list of measuring points.
Stop measured value
-20.0 ... 150.0 °C, [ off ]
(Titrando only)
Stop when the preset measured value has been reached since the start of
the measurement.
off
means no stop.
Evaluations
Command name
25 characters, [ MEAS T # ]
Name of the command.
The following methods for evaluating measuring curves are defined on this tab.
Fix end point evaluation
on, [ off ]
To a fixed quantity (measured value or time) the associated values for the
other quantities are interpolated from the list of measuring points for the
fixed end point.
Fixed quantity
[ Measured value ], Time
Selection of the fixed quantity to which the associated values for the other
quantities are to be interpolated from the list of measuring points.
Fix EP# at
-20.0 ... 150.0 °C, [ off ]
Fixed end point 1 ... 9 for
Fixed quantity
=
Measured value
.