1.4.8 Remote Communications

RS232C Serial Port . . . . . . . . . . . . . Standard; INFICON binary protocol

Baud Rates . . . . . . . . . . . . . . . . . . . 115,200, 57,600, 38,400, 19,200, 9,600

Ethernet . . . . . . . . . . . . . . . . . . . . . . Optional, Programmable IP Address and

Net Mask

USB port. . . . . . . . . . . . . . . . . . . . . . Print screen bitmap (.bmp) files, Datalogging

file, Configuration file store/retrieve

1.4.9 Accessories

Connector Kit . . . . . . . . . . . . . . . . . . Connectors for inputs and relays

Operating Manual. . . . . . . . . . . . . . . 074-5000-G1 CD-ROM

1.4.10 Power

Power . . . . . . . . . . . . . . . . . . . . . . . . 100-230 V (ac) +/-15% V (ac); 50/60 Hz

+/-3Hz Maximum apparent power 150 VA

Fuse . . . . . . . . . . . . . . . . . . . . . . . . . use 250 V 4 Amp, 5 x 20, Slow Blow

1.4.11 Operating Environment

Usage. . . . . . . . . . . . . . . . . . . . . . . . Indoor only

Temperature . . . . . . . . . . . . . . . . . . . 0 to 50°C (32-122°F)

Humidity . . . . . . . . . . . . . . . . . . . . . . Up to 85% RH, non-condensing

Altitude . . . . . . . . . . . . . . . . . . . . . . . Up to 2000 meters

Installation (Overvoltage) . . . . . . . . . Category II per IEC 60664

Pollution Degree. . . . . . . . . . . . . . . . 2 per EN 61010

1.4.12 Storage Temperature

Storage Temperature . . . . . . . . . . . . -10 to 60°C (14 to 140°F)

1.4.13 Warm Up Period

Warm Up Period. . . . . . . . . . . . . . . . None required;

For maximum stability allow 5 minutes.