INFICON IC6 Thin Film Deposition Controller User Manual
Page 278
15 - 8
PN
07
4-
50
5-
P1
F
IC6 Operating Manual
CrossTalk > 100%
One or both of the co-deposited
layers has a Cross Talk
percentage ≥100%.
When starting a co-dep layer the
Cross Talk % is checked. If the
calculated Cross Talk % is
≥100% then there must be a
problem with the cal thickness
values the user entered and the
layer stops with this message as
the cause of stop. If a layer is
running and the cal thickness
values are changed to cause the
calculated Cross Talk % ≥100%
then the layer stops with this
message as the cause of stop
A RESET command
received from the keyboard,
digital input, remote
communications or a Logic
action.
A Start, if the condition
causing the stop has been
cleared and the process can
resume.
All
14
DAC Failure
Monitoring of the DAC output via
an ADC determined that the
output voltage is incorrect.
All
15
Half Rate Fail
AutoTune is unable to vary the
source power to obtain half the
desired deposition rate.
AutoTune
16
Low Rate
Auto tuning is terminated when
the deposition rate at the start of
tuning is below 0.1 Å/s.
AutoTune
17
AutoTune Fail
The complete tune partition of
AutoTune did not determine
reliable system response results.
AutoTune
18
Quick Tune Fail
The Quick tune portion of
AutoTune did not get repeatable
system response results.
AutoTune
19
Bad Co-Dep Load
The loading of co deposition
layers is bad
All
20
Uncalibrated
The co-deposition layers are un
calibrated.
All
21
Bad Layer Load
There was a problem loading a
layer.
All
22
Table 15-4 Stop messages (continued)
Cause of STOP
Message
Set
Clear
Screen
Remote
Comm.
Encodes