Page 278

background image

15 - 8

PN

07

4-

50

5-

P1

F

IC6 Operating Manual

CrossTalk > 100%

One or both of the co-deposited
layers has a Cross Talk
percentage ≥100%.
When starting a co-dep layer the
Cross Talk % is checked. If the
calculated Cross Talk % is
≥100% then there must be a
problem with the cal thickness
values the user entered and the
layer stops with this message as
the cause of stop. If a layer is
running and the cal thickness
values are changed to cause the
calculated Cross Talk % ≥100%
then the layer stops with this
message as the cause of stop

A RESET command
received from the keyboard,
digital input, remote
communications or a Logic
action.
A Start, if the condition
causing the stop has been
cleared and the process can
resume.

All

14

DAC Failure

Monitoring of the DAC output via
an ADC determined that the
output voltage is incorrect.

All

15

Half Rate Fail

AutoTune is unable to vary the
source power to obtain half the
desired deposition rate.

AutoTune

16

Low Rate

Auto tuning is terminated when
the deposition rate at the start of
tuning is below 0.1 Å/s.

AutoTune

17

AutoTune Fail

The complete tune partition of
AutoTune did not determine
reliable system response results.

AutoTune

18

Quick Tune Fail

The Quick tune portion of
AutoTune did not get repeatable
system response results.

AutoTune

19

Bad Co-Dep Load

The loading of co deposition
layers is bad

All

20

Uncalibrated

The co-deposition layers are un
calibrated.

All

21

Bad Layer Load

There was a problem loading a
layer.

All

22

Table 15-4 Stop messages (continued)

Cause of STOP
Message

Set

Clear

Screen

Remote

Comm.

Encodes