10 - 30

PN

07

4-

50

5-

P1

F

IC6 Operating Manual

50

(0x32)

Sensor 1

Failure Action

0/ 4/ 2

Encode

0 = Unused

1 = PostDp

2 = Stop

3 = Suspnd

4 = TimePw

Visible only if Multipoint is

set to Yes

At least one must be

non-zero (when multipoint

set to "Yes")

Update not allowed when

transfer or backup sensor

is running.

51

(0x33)

Sensor 2

Failure Action

0/ 4/ 0

52

(0x34)

Sensor 3

Failure Action

53

(0x35)

Sensor 4

Failure Action

54

(0x36)

Sensor 5

Failure Action

55

(0x37)

Sensor 6

Failure Action

56

(0x38)

Sensor 7

Failure Action

57

(0x39)

Sensor 8

Failure Action

58

(0x3a)

Sensor 1

Failure Trigger

0/ 1/ 1

Encode

0 = OnFail

1 = Last

Visible only if Multipoint is

set to Yes

Gray font if Failure

Action= 0 (Not Used)

Update not allowed when

transfer or backup sensor

is running.

59

(0x3b)

Sensor 2

Failure Trigger

60

(0x3c)

Sensor 3

Failure Trigger

61

(0x3d)

Sensor 4

Failure Trigger

62

(0x3e)

Sensor 5

Failure Trigger

63

(0x3f)

Sensor 6

Failure Trigger

64

(0x40)

Sensor 7

Failure Trigger

65

(0x41)

Sensor 8

Failure Trigger

Table 10-9 Material parameters (continued)

QM UM
Cmd ID

Name

(Units)

Range:

Low Limit/

High Limit/

Default

Data Type

Notes

Update Restrictions

and Display Notes