10.4.15 IC6 Action List

Actions are triggered in the Logic Statement THEN section. Logic codes are used
in the remote communication query and update commands. Some actions that
require a numeric may act on all items by entering a zero, as noted in

Table 10-8

.

Test

0x7D

0x83

AutoTune

0x7E

0x82

Rate Dev(iation)

Err(or)

0x7F

0x81

Note 1: Seconds are entered via remote communications as tenths of seconds. Seconds are
entered by keyboard as whole numbers.
Note 2: The time is entered via remote communications as whole minutes. 99 hours = 5940
minutes, so 99 hours + 59 minutes = 5999 minutes maximum. The time entered by keyboard is in
hours and minutes.
Note 3: Maximum number of minutes is 59. The entry of 00:65 is invalid. That time can be entered
as 01:05.
Note 4: Thick Sum will be sent as an integer Angstroms (the number multiplied by 1000).

Table 10-7 Event list (continued)

Events

1st

numeric

(1 Byte)

(XXX)

Min Max

2nd

numeric

(Integer)

(YYY)

Min

Max

Logic

Code

Logic

Code

Negated

Table 10-8 Action list

Actions

Numeric

(Byte)

Min Max Logic Code

Extern Out On XXX

Output #

1

38

0x41

Extern Out Off XXX

Output #

1

38

0x42

Select Process XXX

Process #

1

50

0x43

Stop

0x44

Start

0x45

Reset

0x46

Suspend XXX

Material #

0

32

0x47

Zero Deposit Time XXX

0x48

Zero Thick XXX

0x49

Trig Fnl Thick XXX

0x4A

Continue Dep XXX

0X4B

Start Deposit XXX

0x4C

Go to NonDep XXX

0x4D

Go to PostDep XXX

0x4E