10.4.15 IC6 Action List
Actions are triggered in the Logic Statement THEN section. Logic codes are used
in the remote communication query and update commands. Some actions that
require a numeric may act on all items by entering a zero, as noted in
.
Test
0x7D
0x83
AutoTune
0x7E
0x82
Rate Dev(iation)
Err(or)
0x7F
0x81
Note 1: Seconds are entered via remote communications as tenths of seconds. Seconds are
entered by keyboard as whole numbers.
Note 2: The time is entered via remote communications as whole minutes. 99 hours = 5940
minutes, so 99 hours + 59 minutes = 5999 minutes maximum. The time entered by keyboard is in
hours and minutes.
Note 3: Maximum number of minutes is 59. The entry of 00:65 is invalid. That time can be entered
as 01:05.
Note 4: Thick Sum will be sent as an integer Angstroms (the number multiplied by 1000).
Table 10-7 Event list (continued)
Events
1st
numeric
(1 Byte)
(XXX)
Min Max
2nd
numeric
(Integer)
(YYY)
Min
Max
Logic
Code
Logic
Code
Negated
Table 10-8 Action list
Actions
Numeric
(Byte)
Min Max Logic Code
Extern Out On XXX
Output #
1
38
0x41
Extern Out Off XXX
Output #
1
38
0x42
Select Process XXX
Process #
1
50
0x43
Stop
0x44
Start
0x45
Reset
0x46
Suspend XXX
Material #
0
32
0x47
Zero Deposit Time XXX
0x48
Zero Thick XXX
0x49
Trig Fnl Thick XXX
0x4A
Continue Dep XXX
0X4B
Start Deposit XXX
0x4C
Go to NonDep XXX
0x4D
Go to PostDep XXX
0x4E